60 Mega-800 SCINCO
6.4.3 Set Protein factors
In the case of custom defined method, press to set protein factors, as shown in Fig. 6.9:
Fig. 6.9 Protein factors setting
At the bottom of this page, enter the first protein factor by pressing the numeric keys ― and
decimal point . The key is used to clear off the preceding entry. Press key to confirm
the numeric entry. The system will prompt the user to input the other protein factor after entering the
first one. Press key to confirm the entry and quit. If nothing is to be entered, press
key to skip the wavelength entering, and the system will adopt the current setting.
6.4.4 Set background correction at 320nm
In the case of custom defined method, press key to set background correction at 320nm. Two
choices are available for this item: YES and NO, which can be selected alternately each time
key is pressed, as shown in Fig. 6.10:
Note:
The input protein factors should be integers or decimals. The range is from -9999 to 9999. If the input
value exceeds 7 digits (including decimal point) or exceeds the range, the system will delete what has been
entered automatically and you should start a new entry again.