entered, press key to return to parameters setting page.
② Set the intercept of calibration curve
On the parameters setting page of K-Factor method, press key to set intercept B, as
shown in Fig. 5.11.
Fig. 5.11 Set the intercept of K-Factor method
At the bottom of this page, enter the intercept B by pressing the numeric keys ― .
The key is used to clear off the entry. Press key to confirm. If nothing is to be
entered, press key to return to parameters setting page.
5.3.2 Standards method
Standards method refers to the method of constructing a calibration curve by using the standards
concentration values and absorbance values.
First, make sure that the measurement mode is ―Standard‖. Press key in the case of
quantitation parameters setting page to gain access to parameters setting page of Standards method,
as shown in Fig. 5.12.
Fig. 5.12 Parameters setting of standards method
Note:
The system accepts integral and decimal, the range of which is –999-999. If the input value
exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system
will delete what has been entered automatically and you should start a new entry again.
Note:
The system accepts integral and decimal, the range of which is –999-999. If the input value
exceeds 7 digits (including decimal point and negative sign) or exceeds the range, the system
will delete what has been entered automatically and you should start a new entry again.