61 Mega-800 SCINCO
Fig. 6.10 Background correction at 320nm
6.4.5 Set calibration factors
In the case of custom defined method, press key to set calibration factors, as shown in Fig.
6.11:
Fig. 6.11 Set calibration factors
At the bottom of this page, enter the calibration factors by pressing the numeric keys and
decimal point . The key is used to clear off the preceding entry. Press key to confirm
the numeric entry. The system will display the factors and return to the parameters setting page if the
entered value meets the requirements. If nothing is to be entered, press key to skip the factor
entering, and the system will adopt the current setting.
6.5 Auto zero
On the main menu of DNA/Protein analysis, press key to set Abs zero automatically at
current measuring wavelength. Before starting auto zeroing, put the blank sample into the cell.
6.6 Measurement
Only one cell can be used for DNA/Protein analysis (No.1 cell is used in the case of multi-cell holder).
On the main menu of DNA/Protein analysis, press key to start measurement. Each time
you press key, it remeasures the sample and displays the resuluts, as shown in Fig. 6.12:
Note:
The input calibration factors should be integers or decimals. The range is from -999 to 999. If the input
value exceeds 7 digits (including decimal point) or exceeds the range, the system will delete what has been
entered automatically and you should start a new entry again.