46 Mega-800 SCINCO
On the parameters setting page, press to set standards number, and press to set the
concentration and absorbance value of a specific standard sample. Press key toto construct
and display the calibration curve.
Set standards number
Press key to set standards number, as shown in Fig. 5.13:
Fig. 5.13 Set standard number
At the bottom of this page, enter standard number by pressing the numeric keys
. The key is used to clear off the entry. Press key to confirm. If
nothing is to be entered, press key to return to parameters setting page.
Set concentration value and absorbance value
On the parameters setting page of standards method, press key to set the specific
standard sample, as shown in 5.14. At the bottom of this page, enter standard No. by
pressing the numeric keys . The key is used to clear off the entry.
Press key to confirm. If the entered number is eligible, the system will prompt the
user to input standard concentration. If nothing is to be entered, press key to
skip it. The default standard number will be used (displayed on the screen).
At the bottom of this page, enter the concentration value by pressing the numeric keys
. The key is used to clear off the entry. Press key to confirm. If
the entered value is eligible, the system will prompt the user to input standard
absorbance value. If nothing is to be entered, press key to skip it. The default
concentration will be used (displayed on the screen).
At the bottom of this page, enter the absorbance value by pressing the numeric keys
. The key is used to clear off the entry. Press key to confirm. If
the entered value is eligible, the system will prompt the user with the information ―Please
select item‖. If nothing is to be entered, press key to quit the standards setting
Note:
The system accepts integral
the range of which is 1-10. If the input value exceeds 7
digits (including decimal point and negative sign) or exceeds the range, the system will
delete what has been entered automatically and you should start a new entry again.