5 – Installation and set-up MultiExact Gas Analyser
6. Ensure that the slide opening is at the front, then loosely fit the right-hand outer
slide section (5) to the front and rear slide support brackets (14, 10) and secure
with the four M4 screws (4), and the nuts and washers (9, 13).
7. Ensure that the front of the right-hand outer slide section (5) is 35 mm behind
the rack enclosure front flange (16), then fully tighten the nuts (9, 13) to secure
the slide section in position.
8. Use the procedure in Steps 4 to 7 to fit the left-hand support clamps, slide
support brackets and outer slide section.
9. Align the ends of the left-hand and right-hand slide inner sections (2) in the
openings in the front of the left-hand and right-hand slide outer sections (5) and
slide the analyser (1) into the rack enclosure.
10. Use the four M6 pan head screws (19) and plastic cup washers (18) to secure
the analyser in place.