MultiExact Gas Analyser 5 – Installation and set-up
Specifies the flushing time (0 min 30 sec to 59 min
59 sec)
On (active) or off (inactive). If active the calibration or
validation can be carried out regularly at a specified
frequency.
Only applicable if the timer is active. This specifies the
time/date on which the first automatic validation/
calibration will be carried out. See Section 5.4.3.
Only applicable if the timer is active. If set (in the range
1
hour to 999 days), specifies the interval between
successive automatic calibrations/validations. Select
0 hours and 0 days if repeat is not required.
Figure 28 shows a typical automatic calibration sequence, where:
• Pre-warning is on.
• Automatic calibration of transducer 1 (Tx1) has been initiated.
• validation/calibration type has been set to low & high.
• Mode has been set to calibration.
• Inerting is set to on (active) and inert gas has been set to low.