I Reducing the Carrier Gas Flow Rate After Sequential Analysis and Using the Ecology Mode
108 GCMS-QP2010 Series Operation Guide
5
Edit the batch file.
Add a row at the end and select a method file created in "Appendix I.1.2 Creating Batch Files" P.106.
Vial number, level number, and injection volume settings do not need to be changed from their default
values. Enter a data file name that is not the same as any other row.
6
Click the [Run Mode] cell for the row that specifies the method file that reduces the flow
rate, then click the arrow button that appears.
The [Run Mode] window opens.
7
Configure [Run Mode] settings as shown below, then click [OK].
8
Name and save the batch file, then click the [Start] icon on the [Batch] assistant bar.
This results in loading the method file that reduces flow rate after the analysis for the seventh row is
finished, which ends the sequential analysis with the carrier gas flow rate at 20 mL/min.