TROUBLESHOOTING AND CLEARING MALFUNCTIONS
10.3 Integrated monitoring and diagnosis system
The GM35 is equipped with an integrated system that constantly monitors the operating
state of the SR-unit and evaluation unit. Appropriate messages are generated and logged
in the devices for subsequent evaluation should any deviations from normal operating
states occur.
Messages for the two system components are categorized into error messages and
warning messages depending on the anticipated effects:
• Warning messages are generated if the measurement results are not (yet) directly
affected by the change in the system state. Nevertheless, it is important that the cause(s)
be investigated and corrected, e.g. by means of maintenance measures, to avoid further
malfunctions and damage to the device in particular.
• Error messages are generated when measuring operation is no longer possible or no
longer reliable.
These warning and error messages are stored in the integrated message memory of the
evaluation unit.
Message signaling and retrieval on the evaluation unit
* The relay is active during normal operation (no malfunction), i.e. the contact is closed.
Procedure
Troubleshooting Table If a warning or malfunction is signaled, first call up pending error messages in the Error
menu. Then localize the possible cause and clear the malfunction; see Troubleshooting
Table.
maintcalpardiag
GM 35
Evaluation Unit
CO
CO
2
H
2
O
mg/
m
3
236
Ref. conditions
Hum: wet
Measuring
Operation
Service
Warnin
Malfunction
Component/Tool Signal Remark
Front panel
Evaluation unit
Warning LED goes on
Functional impairment on system that
will not directly lead to corrupt
measured values.
Malfunction LED
blinks
Functional impairment on system that
can lead to system failure or
restricted function. See error memory,
logbook.
Diagnosis mode
Error memory ▸ Call up menu Error
Plain-text message(s) for errors that
have occurred to localize and clear
the problem.
See “Troubleshooting Table”.
Warning memory
▸ Call up menu Warn-
ing
Plain-text message(s) for pending
warnings.
Output
for problems
(malfunction, error
message)
Relay 1 inactive* Group malfunction.
Operation
Service
Warning
Malfunction
Error indication Possible cause Clearance
● LED Malfunction blinks;
(LED Warning could go
on)
● Relay 1:
Group malfunction
Plain-text error messages
indicate possible causes
▸ Start Diagnosis mode (diag):
–Call up menu Error (or Warning)
– Check and clear the specified malfunction.