START-UP
Fig. 52: Connecting the power supply to the SR-unit for adjusting the measuring system
7.5.3 Aligning the optical axis of the measuring probe
To ensure correct measuring function, the optical axis of the measuring probe must be
aligned exactly to the light beam of the SR-unit. To do this, adjust the 2 screws on the
device flange,
see page 111 below.
2 options are available for checking the probe position in XY direction:
On the visor on the right side of the SR-unit enclosure,
see page 111
On the EvU display after switching to Maintenance mode (maint) and calling up menu
ADJ. OPT ALIGN, see page 86.
This menu item must always be carried out when the “Automatic Alignment Control”
option is implemented in the GM35 SR-unit.
Checking on the visor
Check the alignment of the optical axis using the visor on the right-hand side of the SR-unit
enclosure and by making adjustments on the probe fixing on the device flange.
Power supply cable with round plug-in
connector, 2017519
CAN bus cable to evaluation
unit (possibly via terminal box)