8009389/YN39/V3-0/2015-08| SICKOPERATING INSTRUCTIONS|GM35
Subject to change without notice
TROUBLESHOOTING AND CLEARING MALFUNCTIONS
10.3.5 Warning messages for the measuring probes
10.3.6 Further tips on troubleshooting
Troubleshooting on the evaluation unit
Fig. 69: LED displays, signals and fuses on the evaluation unit
Service requirement–
message on the EvU
Component/possible causes Clearance
El too hot Electronics too hot. Improve GPP cooling.
Heater defect Heater defective. Repair heater.
Heating too low
Setpoint temperature not
reached.
Set setpoint temperature
lower.
Contact Service.
I_Heat < 1.5 A
Heating current too low
(GPP).
Check heating control cable,
repair when necessary.
I_Heat no signal No heating current (GPP). Check heating control.
P no signal
No signal from pressure
sensor.
Check pressure sensor cable,
repair when necessary.
P out of range Pressure sensor without
valid measured value.
Check pressure sensor, repair
when necessary.
[ t ] no signal No signal from temperature
sensor.
Check temperature sensor
cable, repair when necessary.
Relay outputs 1...4
Analog output
Fuses
2.5 AT, 250 V
Operating voltage:
230 V AC – no jumper
115 V AC – jumper
5 V DC
LEDs light up:
Internal voltage supply
present
24 V DC
LEDs light up when
inputs are connected,
i.e. a signal is available.
Digital inputs 1, 2 and 3