8009389/YN39/V3-0/2015-08| SICK OPERATING INSTRUCTIONS|GM35
Subject to change without notice
TROUBLESHOOTING AND CLEARING MALFUNCTIONS
H2O Reference
Deviating control value
measurement too high.
Carry out zero adjust on
GM35;
see page 87.
Perform maintenance,
see
page 99
.
Contact Service if necessary.
HYGRO com
No connection between CO
2
/H
2
O
module and humidity sensor.
Check cable connection is
secure with correct seat;
repair if necessary.
Switch to default value.
If the error cannot be cleared:
Contact Service.
HYGRO com.
“Hygro” communication
interrupted.
Sensor defective or cable
broken.
Contact problem.
System switches to default
value
Check cable connections.
Exchange sensor.
HYGRO internal
No communication with internal
humidity sensor.
Sensor defective.
Cable broken.
Check cable connections.
Exchange sensor.
MIRROR adj.
Automatic beam tracking:
Deviation of optical axis is
greater than area that can be
tracked.
Realign optical axis between
SR-unit and reflector,
see
page 111
.
MR Adjustment
Visor unit shows deviation
(> 0.5 of setpoint position).
Realign optical axis between
SR-unit and reflector;
see
page 111
.
Software version EvU and SR-unit incompatible.
Install compatible (current)
software for EvU and SR-unit.
Contact Service if necessary.
VISOR amplifier Amplifier in visor at limit.
Check contamination; clean
optical interfaces,
(page 104f).
Exchange IR source or contact
Service.
Warning message Component/possible causes Clearance