Operating Instructions 7-21 M90 Fusion Splicer
Issue 02/97 Troubleshooting and Care/Maintenance
Fault Message
Fault Symptom
Possible Cause Fault Locating Remedy
Fault messages
relating to the
splice memory
No data avail-
able
The splice mem-
ory is empty.
Ensure that the splice
memory is switched on
(automatic or manual) in
the parameters.
Attenuation value
flashing
The attenuation
value of the
splice could not
be determined
reliably due to a
fault (
end face
dirty or dam-
aged; bad
cleavage
).
See appropriate fault mes-
sages.
Display of
Fault
instead of at-
tenuation value
The attenuation
could not be
calculated for the
splice concerned
(
attenuation
measurement
not possible
or
bad cleavage
).
See appropriate fault mes-
sages.
Fault messages
during printing
Printer not
ready!
The printer con-
nected to the
serial interface
transmits incor-
rect characters.
Check the baud rate set at the printer and
splicer.
Ensure that the printer supports the XON/XOFF
protocol and the data format (8 data bits, 1 stop
bit, no parity).