Functions
6.4 [2.03] Settings > Measuring ranges
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
103
4. Enter the current time: "Time" parameter field.
5. Enable/disable automatic switching to daylight saving time: "Daylight saving time"
parameter field.
Select one of the alternatives:
Automatic switching is disabled.
Autom. EU Automatic switching to CEST ("Central European Summer Time") on the last
Sunday in March. The CEST ends on the last Sunday in October.
Autom. US Switching to U.S. DST ("Daylight Saving Time") on the second Sunday in March.
DST ends on the first Sunday in November.
[2.03] Settings > Measuring ranges
6.4.1
Overview of measuring ranges
Measuring ranges
Note
Putting back into operation with disabled measuring ranges
The measuring ranges are activated in the factory. If you disable the measuring ranges
(setting: "Off"), all the setting
options in other menus related to the measuring range are also
hidden. The measuring ranges can no longer be used.
If you put the device back into operation with fully deactivated measuring ranges, the status
bar of the display shows message 150 "Paramete
r value invalid". Before you continue to put
the device back into operation:
Activate at least one measuring range
Disable the autoranging function if it is enabled To use the automatic measuring range
switching, at least two measuring ranges must be active.