Functions
6.1 [1] Quick Start
Operating with the Local User Interface
96 Operating Manual, 06/2017, A5E31930478-05
1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
2. Call autoranging: Navigation line "Autoranging".
The parameter display of autoranging has a parameter field and a display box.
3. Specify measuring range: "Measuring range selection" parameter field.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, adjust the start and end values for the individual
measuring ranges: [2.08] Settings > Calibrations (Page 114).
6.1.7.2
Execute calibration/validation
Procedure
1. Component > Main menu > "1 Quick Start" > "5. Calibrations"
Depending on the wizard you enable, you either execute a calibration or validation. The
parameter assignment of this menu is used for both processes.
2. Select components to be calibrated/validated: "Components" parameter field.
– Authorize access with Standard PIN, if necessary.
– Enable/disable one or more component-related check boxes in the displayed editing
window.
3. Specify calibration type: "Calibration type" parameter field.
– If you select "2-pt.", zero gas calibration and span gas calibration are executed one
after the other. Otherwise you set either a zero gas calibration or a span gas
calibration.
4. Specify the scope of the calibration: "Single/total" parameter field.
By setting this parameter you define the scope of the calibration - one or all measuring
ranges. The measuring range set in the calibration is always calibrated:
– If you set "Single", calibration is only applied for the set measuring range. Calibrations
of the other measuring ranges remain unchanged.
– If you set "Total", the calibrations of the other measuring ranges are overwritten by the
calibration of this "leading" measuring range.
In validation, only the set, "leading" measuring range is considered.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals