Functions
6.1 [1] Quick Start
Operating with the Local User Interface
96 Operating Manual, 06/2017, A5E31930478-05
1. "1 Quick Start" > "4. Measuring ranges" > "n component" main menu
The menu includes four navigation lines for setting the measuring ranges and one
navigation line for setting autoranging.
2. Call autoranging: Navigation line "Autoranging".
The parameter display of autoranging has a parameter field and a display box.
3. Specify measuring range: "Measuring range selection" parameter field.
– Authorize access with Standard PIN, if necessary.
– Select one of the offered measuring ranges or autoranging:
If the message "Active" is displayed in the "Autoranging" line, autoranging is activated.
If "Not possible" is displayed, adjust the start and end values for the individual
measuring ranges: [2.08] Settings > Calibrations (Page 114).
Execute calibration/validation
Procedure
1. Component > Main menu > "1 Quick Start" > "5. Calibrations"
Depending on the wizard you enable, you either execute a calibration or validation. The
parameter assignment of this menu is used for both processes.
2. Select components to be calibrated/validated: "Components" parameter field.
– Authorize access with Standard PIN, if necessary.
– Enable/disable one or more component-related check boxes in the displayed editing
window.
3. Specify calibration type: "Calibration type" parameter field.
– If you select "2-pt.", zero gas calibration and span gas calibration are executed one
after the other. Otherwise you set either a zero gas calibration or a span gas
calibration.
4. Specify the scope of the calibration: "Single/total" parameter field.
By setting this parameter you define the scope of the calibration - one or all measuring
ranges. The measuring range set in the calibration is always calibrated:
– If you set "Single", calibration is only applied for the set measuring range. Calibrations
of the other measuring ranges remain unchanged.
– If you set "Total", the calibrations of the other measuring ranges are overwritten by the
calibration of this "leading" measuring range.
In validation, only the set, "leading" measuring range is considered.