6.5 [2.04] Settings > Limits
Operating with the Local User Interface
108 Operating Manual, 06/2017, A5E31930478-05
Monitoring options
The following monitoring options can be used:
Table 6- 2 Monitoring options using limits
Monitoring
Setting
None The low and high limits are inactive: Setting "Off"
for both.
Low limit violated
Only the low limit is activated.
High limit exceeded
Only the high limit is activated.
LImit violated at "Value outside" The low and high limits are activated. Limit viola-
tion: The measured value is located outside the
interval.
LImit violated at "Value within" The low and high limits are activated. Limit viola-
tion: The measured value is located within the
interval.
See also
Setting limits (Page 108)
6.5.2
Setting limits
Procedure
Note
Effects
Changing the parameter setting has a direct effect on the following
functions:
Status bar: Display of system messages
Logbook: entries in the device logbook
Digital measured value status
Note
Component-specific setup
Limit ranges are component
-specific. Repeat the procedure described below for all
components, if necess
ary.