Functions
6.16 [3.02] Maintenance & Diagnostics > Messages to be acknowledged
Operating with the Local User Interface
206 Operating Manual, 06/2017, A5E31930478-05
[3.02] Maintenance & Diagnostics > Messages to be acknowledged
6.16.1
Displaying messages to be acknowledged
Procedure
Note
Distinction
SIPROCESS
GA700 distinguishes between messages that require acknowledgment and
messages that do not require acknowledgment.
Messages not requiring acknowledgement are cleared if the problem causing them no longer
exists. Messages
that require acknowledgment are identified by the symbol â–¡ and are
revoked when the cause of the message is no longer present and an acknowledgment is
made.
Note
Device behavior after voltage off/on
After the power supply has been switched off and on, messages that require
acknowledgment are treated as follo
ws:
Analyzer module(s): Messages that have not been acknowledged but require
acknowledgement are still pending.
Device: Messages that have not been acknowledged but require acknowledgement were
deleted.
For additional information, see also →
OXYMAT 7 message list (Page 257).
1. Main menu > "3. Maintenance & Diagnostics" > "2 Messages to be acknowledged"
The menu contains the following entries:
Structure of the navigation lines
Messages to be acknowledged
2. → Acknowledge message (Page 207). All messages to be acknowledged are listed
unfiltered.
3. → Acknowledge all messages (Page 208). All messages requiring acknowledgment can
be acknowledged in total with this function.