Functions
6.19 [3.05] Maintenance & Diagnostics > Diagnostics values
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
215
[3.05] Maintenance & Diagnostics > Diagnostics values
6.19.1
[3.05.1] Measured signal diagnostics
6.19.1.1
This menu provides you access to status information/current values that you can use to
evaluate the quality of the measured signal:
The access paths vary depending on the installed analyzer modules:
● OXYMAT 7/CALOMAT 7:
Component > "3. Maintenance & Diagnostics" > "05. Diagnostics values" > "1. Measured
signal diagnostics" > "1. Measured signal".
● ULTRAMAT 7:
– Component > "3. Maintenance & Diagnostics" > "05. Diagnostics values" > "1.
Measured signal diagnostics" > "1. Measured signal".
– Component > "3. Maintenance & Diagnostics" > "05. Diagnostics values" > "1.
Measured signal diagnostics" > "1. Measured signal Part 2".
You only have read access for the majority of parameters in these menus.
OXYMAT 7: Evaluation of the measured signal quality
Table 6- 52 OXYMAT 7: [3.05.1.1] Measured signal
Activation/deactivation by service personnel
Vibration compensation Only f
or devices with vibration compensation. Can be enabled/disabled with
Expert PIN.
Only for devices with vibration compensation.
Sensor signal converted to mV
Current temperature value