Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
264 Operating Manual, 06/2017, A5E31930478-05
7.3
ULTRAMAT 7 message list
Mes-
sage
Cause
Remedy
No.
Text
2 Maximum warming-up
time exceeded.
Device is operated outside the permit-
ted ambient temperature range
Check the ambient temperature of the
analyzer
Check if there are additional messag-
es and follow the instructions.
3 Temperature on measur-
ing chamber too high.
Device is operated outside the permit-
ted ambient temperature range
Errors in temperature control
Check the ambient temperature of the
analyzer
Check the temperatures of the meas-
uring chamber and analyzer module
in the menu "[3.05.1.3] Heating con-
troller".
4 Temperature on measur-
ing chamber too low
Temperature sensor analyzer cham-
ber short-circuited / defect
Errors in temperature control
Check the ambient temperature of the
analyzer
Check the temperatures of the meas-
uring chamber and analyzer module
in the menu "[3.05.1.3] Heating con-
troller".
7 Temperature on chopper
too high.
Device is operated outside the permit-
ted ambient temperature range
Errors in temperature measurement
Check the ambient temperature of the
analyzer and in the menu "[3.05.1.1]
Measured signal" the temperatures of
the chopper and analyzer module.
8 Temperature on chopper
too low.
Device is operated outside the permit-
ted ambient temperature range
Errors in temperature measurement
Check the ambient temperature of the
analyzer
Check the temperatures of the chop-
per and analyzer module in the menu
"[3.05.1.1] Measured signal".
9
Temperature on probe too
high.
Device is operated outside the permit-
ted ambient temperature range
Error in the temperature measure-
ment
Check the ambient temperature of the
analyzer
Check the temperatures of the meas-
uring head/detector and analyzer
module in the menu "[3.05.1.1] Meas-
ured signal".
10 Temperature on probe too
low.
Device is operated outside the permit-
ted ambient temperature range
Error in the temperature measure-
ment
Check the ambient temperature of the
analyzer
Check the temperatures of the meas-
uring head/detector and analyzer
module in the menu "[3.05.1.1] Meas-
ured signal".