Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
265
Mes-
sage
Cause
Remedy
No.
Text
11 Maximum temperature
exceeded. Heaters turned
off.
Safety function
Ambient temperature too high
One or more heater control loops
faulty
Wait until the device has cooled
down.
Then acknowledge the message.
You should a
lso note the messages in
the logbook.
14 Atmospheric pressure
measurement outside
tolerance.
Atmospheric pressure< 500 hPa or >
1500 hPa
Atmospheric pressure sensor faulty
Hose connection to atm. pressure
sensor loosened
Check the input of the atmospheric
pressure sensor.
15 Sample gas pressure
outside tolerance.
Pressure sample gas < 500 hPa or >
1500 hPa
Sample gas outlet closed
pressure sensor faulty
Gas path to the pressure sensor un-
tight
Gas path to the pressure sensor
blocked
Check the sample gas pressure in the
menu "[2.20.04] Calibrate pressure
sensor".
20 Intensity outside toler-
ance.
Source defective
Receiver chamber faulty
Exchange the IR radiators.
Clean the cuvette.
22 Calibration tolerance
exceeded at zero point.
Zero gas changed, setpoint faulty
Drift in the analyzer
Check the setpoint for the zero gas
calibration in the menu "[2.08.1.1]
Setpoints"
O7/U7: Check the reference gas
supply
23 Validation tolerance ex-
ceeded at zero point.
Carry out a zero gas calibration.
24 Calibration tolerance
exceeded at span point.
Span gas changed, setpoint faulty
Drift in the analyzer
Check the setpoint for the span gas
calibration in the menu "[2.08.1.1]
Setpoints"
U7/O7: Check the reference gas
supply.
25 Validation tolerance ex-
ceeded at span point.
Carry out a span gas calibration.
27 Signal frequency
changed. Normalization
required!
Carry out a normalization calibration
menu "[2.20.3.3] Normalization"
28 Control reserve used up.
Check the zero gas and reference
gas supplies.