Alarm, error, and system messages
7.2 OXYMAT 7 message list
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
257
Note
Message behavior at a restart
After a restart, the messag
es with the messages numbers 22 to 32 remain in the message
list. All other previous messages are deleted.
Table 7- 7 OXYMAT 7 message list
2 Maximum warming-up time ex-
ceeded
• Device is operated outside the
permitted ambient temperature
range
• Errors in temperature control
• Check ambient temperature
• Check if there are additional mes-
sages and follow the instructions.
3 Temperature on measuring
chamber too high
• Device is operated outside the
permitted ambient temperature
range
• Errors in temperature control
• Check ambient temperature
• Replace analyzer module
4 Temperature on measuring
chamber too low
•
Temperature sensor of measuring
chamber has short-circuit/is faulty
• Heating controller faulty
• Check ambient temperature
• Replace analyzer module
5 Temperature on heating cartridge
too high
• Temperature sensor of heating
cartridge is not connected/is faulty
• Heating controller faulty
• Check ambient temperature
• Replace analyzer module
6 Temperature on heating cartridge
too low
• Temperature sensor of heating
cartridge is not connected/is faulty
• Heating controller faulty
• Check ambient temperature
• Replace analyzer module
9 Temperature on probe too high
• Device is operated outside the
permitted ambient temperature
range
• Temperature control measuring
head faulty
• Check ambient temperature
• Replace analyzer module
10 Temperature on probe too low
• Device is operated outside the
permitted ambient temperature
range
• Temperature control measuring
head faulty
Check ambient temperature, replace
analyzer module, if necessary