Alarm, error, and system messages
7.2 OXYMAT 7 message list
Operating with the Local User Interface
258 Operating Manual, 06/2017, A5E31930478-05
Message
Cause
Remedy
No.
Text
11 Maximum temperature exceeded.
Heaters turned off automatically.
Safety function
Ambient temperature too high
One or more heater control loops
faulty
Check ambient temperature
Check the other messages includ-
ing the logbook and act accord-
ingly
15 Sample gas pressure outside
tolerance
Pressure sensor faulty
Reference gas open, sample gas
input/output closed
Set sample gas pressure to per-
mitted range.
If sample gas input/output is
closed, the reference gas must be
turned off.
Replace analyzer module
18 Reference gas pressure too low
Reference gas pressure - sample
gas pressure < 2 000 hPa
Sample gas path closed
Reference gas path leaking
Check pressure of reference gas
source
Check flow of reference gas, see
operating instructions.
22 Calibration tolerance exceeded at
zero point
Zero gas changed, setpoint faulty
Incorrect reference gas
Drift probe
Check setpoint
Check reference gas
Replace analyzer module
23 Validation tolerance exceeded at
zero point
Calibration successful Execute zero point calibration
24 Calibration tolerance exceeded at
span point
Span gas changed, setpoint faulty
Incorrect reference gas
Drift probe
Check setpoint
Check reference gas
Replace measuring head
25 Validation tolerance exceeded at
span point
Calibration successful Execute span calibration
27 Signal frequency changed Frequency change executed Normalization required; contact Ser-
vice if necessary
28 Control reserve used up Difference of O
2
concentration be-
tween zero gas and reference gas
too large.
Check zero gas, reference gas
pressure
Carry out zero and full-scale val-
ue calibrations
29 to
32
Zero drift MR1 to MR4 outside
tolerance
Gas line dirty, gas cooler faulty
Gas lines leaking
Incorrect zero gas or setpoint
faulty
Check gas preparation, e.g.,
check filter
Check gas lines for leaks
Check setpoint
33 Measured value greater than full-
scale value
Sample gas concentration too
high
Sample gas pressure too
high/outside pressure correction
range
Check sample gas concentration
Check sample gas pressure
Execute span gas calibration
34 Analyzer module temperature not
in permissible range
Device is operated outside the per-
mitted ambient temperature range
Check ambient temperature