Alarm, error, and system messages
7.2 OXYMAT 7 message list
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
263
Message
Cause
Remedy
No.
Text
237 AutoCal 2 canceled
AutoCal could not be executed due to
a fault.
Ch
eck and remedy pending errors
in the list of current messages.
Then: Execute AutoCal manually.
238 Cross-interference measured
value uncertain or faulty
External or internal cross-
interference
measured value is faulty or the exter-
nal cross-interference measured
value is not transmitted.
Check the sources for the measured
values of the parameterized interfer-
ence gases.
Check source of measured value
for assignable interference gases
Check interference gas supply
If transmission is via MODBUS
TCP, check if errors no. 131 to
137 are pending and follow their
instructions.
239 External pressure value faulty External pressure value is not trans-
mitted or is faulty.
Check external pressure meas-
urement
If transmission is via MODBUS
TCP, check if errors no. 131 to
137 are pending and follow their
instructions.