Functions
6.7 [2.08] Settings > Calibrations
Operating with the Local User Interface
122 Operating Manual, 06/2017, A5E31930478-05
1. Main menu > "2. Settings" > "08. Calibrations" > "1. Setpoints, tolerances, purge
time" > "1. Setpoints"
In this component-specific menu you can enter the setpoints for zero gas and span gas
calibrations for each measuring range.
The unit and format of the input correspond to the default settings of the respective
component.
Structure of the navigation bars
Zero gas measuring range 1 to 4
1)
2)
Span gas measuring range 1 to 4
1)
Setpoint
2)
Reduced display. One line or input of a setpoint is possible for each measuring range.
2)
Entry in the preset unit.
Note
Setting ranges
The setting ranges of the setpoints (possible minimum/m
aximum values) are limited by
-of-scale values and full-scale values. Example:
Changes to a measuring range can result in the setpoints assigned to the measuring
range being outside the measuring range.
The device displays the message "[150] - Parameter value invalid".
Adapt the setpoints. Make sure that the setpoints are within the valid measuring range.
2. Enter zero gas setpoints: .
"Zero gas range 1" to "Zero gas range 4" parameter boxes.
– Authorize access with Expert PIN, if necessary.
– The assignment of one setpoint is sufficient for a total calibration.
– Zero gas setpoint < span gas setpoint
3. Enter span gas setpoints:
"Span gas range 1" to "Span gas range 4" parameter boxes.
– The assignment of one setpoint is sufficient for a total calibration.
– Span gas setpoint > zero gas setpoint
Recommendation: Span gas setpoint at 70 to 100% of the current measuring range.