EasyManua.ls Logo

Siemens SIPROCESS GA700 - Page 162

Siemens SIPROCESS GA700
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Functions
6.9 [2.10] Settings > Correction of cross-interferences
Operating with the Local User Interface
162 Operating Manual, 06/2017, A5E31930478-05
Table 6- 39 OXYMAT 7: Application 3
Description of application 3
Conditions
The actual interference gas concentration is variable.
The influence of interference gas is known and stored in the interference gas reference table.
Sequence
Manual setting of the interference gas component for selection in the interference gas refer-
ence table.
Selection of measured value source for the interference gas. The interference gas is meas-
ured in the selected unit.
Enter substitute value: If the measured value of the interference gas concentration is invalid
or not available, the device works with the concentration of the substitute value.
Linear correction coefficient k1 from device-internal interference gas reference table.
Result The influence of the interference gas is determined from the measured interference gas and the
interference gas reference table stored in the device.
Case 1: The interference gas concentration is measured and is valid.
Case 2: The interference gas concentration cannot be measured or is invalid. A substitute
value is used instead of the measured interference gas value.
Table 6- 40 OXYMAT 7: Application 4
Description of application 4
Conditions
The interference gas concentration is variable.
The influence of interference gas is known but not stored in the interference gas reference
table.
Sequence
Selection of measured value source for the interference gas.
The interference gas is measured in the selected unit here.
Enter substitute value in the selected unit: If the measured value of the interference gas con-
centration is invalid or not available, the device works with the concentration of the substitute
value.
Input of correction coefficient k1 in % always based on 100 % of the interference gas.
Result The influence of the interference gas is determined from the measured interference gas and the
input coefficient.
Case 1: The interference gas concentration is measured and is valid.
Case 2: The interference gas concentration cannot be measured or is invalid. A substitute
value is used instead of the measured interference gas value.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Related product manuals