6.9 [2.10] Settings > Correction of cross-interferences
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
167
Note
Avoid loops (with the variable correction of cross-interference)
Loops in the correction of cross
-references can result in incorrect measured values. The
device does not detect loops. The measured
-value status is set to "Uncertain" or "Invalid".
Assign the parameters for the correction of cross
-references so that the components are not
corrected reciprocally with each ot
her. Avoid the following corrections, for example:
Component 1 corrects component 2. Component 2 corrects component 1.
Note
Activation of correction of cross-interference
In the factory setting, all available process tags are activated. When the measu
ring point
switchover is deactivated, however, the correction of cross
-interference is always active.
With the process tag switchover switched on, the following applies: selection of the active
process tag simultaneously activates the correction of cross
-interference.