EasyManuals Logo

Siemens SIPROCESS GA700 User Manual

Siemens SIPROCESS GA700
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #167 background imageLoading...
Page #167 background image
Functions
6.9 [2.10] Settings > Correction of cross-interferences
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
167
Note
Avoid loops (with the variable correction of cross-interference)
Loops in the correction of cross
-references can result in incorrect measured values. The
device does not detect loops. The measured
-value status is set to "Uncertain" or "Invalid".
Assign the parameters for the correction of cross
-references so that the components are not
corrected reciprocally with each ot
her. Avoid the following corrections, for example:
Component 1 corrects component 2. Component 2 corrects component 1.
Note
Activation of correction of cross-interference
In the factory setting, all available process tags are activated. When the measu
ring point
switchover is deactivated, however, the correction of cross
-interference is always active.
With the process tag switchover switched on, the following applies: selection of the active
process tag simultaneously activates the correction of cross
-interference.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals