General information
2.3 Information on use
Operating with the Local User Interface
18 Operating Manual, 06/2017, A5E31930478-05
If you are using SIPROCESS GA700 devices in critical processes, observe the following
information:
Monitor the limits externally via the control system.
Read out the device status only via the digital outputs of the analyzer module.
Use an available manual measuring range switchover instead of the autoranging.
Only use the "Total" parameter with the "Off" setting for the measuring range switchover.
If you use the "Correction of cross-interference" and "External pressure compensation"
functions, store suitable substitute values.
Only access the measured value via an analog output.