Alarm, error, and system messages
7.4 CALOMAT 7 message list
Operating with the Local User Interface
272 Operating Manual, 06/2017, A5E31930478-05
Message
Cause
Remedy
No.
Text
146 Configuration faulty
• Content of parameter memory faulty
and cannot be restored
• The active device configuration is
faulty
• Load the user or factory data in the
work data.
148 Saved user data faulty
• Content of user data in parameter
memory faulty and cannot be restored
• Save the work data in the user data
again
• Alternatively: Transfer a valid parame-
ter set to the device
149 Factory data faulty
• Content of factory data in parameter
memory faulty and cannot be restored
• The saved factory data is faulty
• Contact Service
150 Parameter value invalid
• Change of input limits of a parameter
due to change of another parameter
• The previously set value is now out-
side the changed input limits
• Due to a change in a parameter, the
value of another parameter is outside
its valid limits
• Correct the value of the associated
parameter
151 Configuration restored
following fault
• Content of parameter memory was
faulty but has been recovered
• An error was detected when loading
the device configuration
• However, a valid configuration could
be recovered without data loss
153 Factory synchronization
of the electronics faulty
• Synchronization values of analyzer
module/option module electronics
faulty and cannot be restored
•
The synchronization data saved at the
factory for the module electronics is
corrupt
• Replace the module
154 Saved sensor module
data faulty
• Content of sensor module data in
parameter memory faulty and cannot
be restored
• The saved sensor data is faulty
• Contact Service
226 Gas warning preliminary
alarm
• Gas warning (preliminary alarm)
227 Gas warning main alarm
• Gas warning alarm (main alarm)
238 Cross-interference meas-
ured value uncertain or
faulty
• External or internal interference gas
measured value is faulty
• Alternatively: the external interference
gas measured value is not transferred
• Check the sources for the measured
values of the set interference gases.
239 External pressure value
faulty
• External pressure measured value is
not being transferred or is faulty
• Check the external pressure meas-
urement

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals