Alarm, error, and system messages
7.5 Wizard-based error messages
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
275
Error ID
Message text (cause/remedy)
F1027
"F1027: The component calibration was not suc-
cessful because the total calibration or another
component calibration has failed.
Check the other component calibration or the
total calibration. This component can only be
successfully calibrated after troubleshooting."
F1029 "F1029: The validation tolerances of the zero gas
and span gas has been violated.
Calibration necessary."
F1030 "F1030: The validation tolerance of the zero gas
has been violated.
Calibration necessary."
F1031 "F1031: The validation tolerance of the span gas
has been violated.
Calibration necessary."