Functions
6.1 [1] Quick Start
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
93
Calibration parameters are assigned manually. Calibrations are distinguished by the
following features:
Person group:
"Customer calibrations" (zero point and/or span gas calibrations)
"Factory calibrations" (normalization and phase calibration).
The section below only covers the area of "Customer calibrations". Information on factory
calibrations is available under → [2.20.03] Factory calibrations. Contact you Siemens
sales rep for more information.
Single/total: Calibration of only one or all measuring ranges
Measuring range
Component
After the start of a calibration, a wizard in the display supports you with preset calibration
steps.
Validation
You use a validation to check whether the deviation from the last calibration is still within the
validation tolerance. If the deviation is outside the validation tolerance, a new calibration is
necessary.
Validation uses the same process steps as calibration. But the validation operation checks
whether the deviation between the actual value and setpoint is still within an assignable
tolerance band. No calibration factors are changed in the process.
If a validation detects a tolerance violation, a corresponding message is output. The
message disappears after a successful calibration.