Page 316 (Model J)
Pass Fail
Step 65:Pressure Support functional test.
Analyser reads PIP as 25mbar
Step 66:Pressure Support functional test.
Ventilator displays PIP as 2-13mbar
Step 70: HFO+CMV functional test.
the ventilator cycles with oscillations in both
the inspiratory and expiratory phases the
waveforms appear in the waveform windows
the I:E ratio in the breath parameter panel
reads 1.0 : 1.0
Step 71: HFO+CMV functional test.
The ventilator cycles with oscillations in the
expiratory phase only.
Step 73: Pressure Change alarm.
Alarm is triggered.