9 Printouts
97
10085896_H
V-PRO
®
maX 2
Operator Manual
========================
======= Not Set ========
========================
=== L E A K T E S T ===
========================
CYCLE START AT 12:11:59
ON 12/01/17
CYCLE COUNT 3
STERILIZER NO 0
SERIAL # Not Set
OPERATOR ______________
REMARKS: _______________
- TIME T=°C P=Torr
------------------------
C 12:11:59 50.2 756.0
L 12:15:55 49.6 0.4
L 12:17:55 50.1 0.4
L 12:27:55 49.6 0.5
========================
Leak Test Result
0.012 Torr/min - Pass
========================
A 12:28:00 49.6 0.5
A 12:34:00 50.0 0.4
Z 12:34:35 50.0 747.3
LOAD 120117-00003
TOTAL CYCLE = 22:36
PRINTOUT CHECKED BY:
________________________
========================
=== L E A K T E S T ===
==== CYCLE COMPLETE ====
====== CYCLE PASS ======
========================
Figure 9-3. Leak Test Printout (Typical)
Printer Self Test
The self-test printout (Figure 9-4) consists of the printer's current Model, H/W Revision, F/W Revision, Memory, RS232
Setting, the Print Mode, the Option Register, and two sample font print outs.