7 — Troubleshooting
73
10085896_H
V-PRO
®
maX 2
Operator Manual
7.58 Too Long to Extend Tube — ALM
Condition
Occurs if LS8 does not sense draw tube extended within two seconds of SV13 activation when Sterilant Cartridge (Cup)
is inserted in cartridge receiver and a cycle is selected.
NOTE:
Draw tube cylinder extend sensor (LS8) is a normally open switch that closes when the draw tube cylinder is fully
extended.
General Troubleshooting
Additional Information
Acknowledge alarm.
1. If alarm clears, run empty chamber Non Lumen
Cycle.
2. If it passes, place unit back into service.
3. If alarm does not clear, contact STERIS.
7.59 Too Long To Fill (Reservoir) — ALM
Condition
Occurs if reservoir fill sensor (LLS1) does not sense H2O2 within 100 seconds from start of phase. Sterilizer
automatically aborts cycle.
NOTE: Reservoir fill sensor (LLS1) is a normally open switch that closes when fluid is detected in
reservoir.
General Troubleshooting Additional Information
Acknowledge Alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
2. If alarm clears and changes to CUP EMPTY alarm,
insert new Cup and place Unit back into service.
3. If alarm does not clear, contact STERIS.