EasyManua.ls Logo

Steris V-PRO maX 2 - Door Issue - Nonalm; Door Open - ALM; Door Open too Long - ALM

Steris V-PRO maX 2
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7 Troubleshooting
59
10085896_H
V-PRO
®
maX 2
Operator Manual
7.21 Door Issue NonALM
Condition
Unit passes cycle and door does not open (Non-Alarm).
General Troubleshooting Additional Information
Vacuum holding door closed.
1. Wait 15 minutes and try to open door again.
2. If door does not open, call STERIS.
7.22 Door Open ALM
Condition
Occurs if door A or B closed switch (LS1/LS2) opens during a cycle. Sterilizer automatically aborts cycle.
NOTE:
Door closed sensors (LS1/LS2) are normally open switches that close when door A/B are closed.
General Troubleshooting Additional Information
Door is not fully closed.
1. Load must be repackaged and reprocessed after
Abort is completed.
2. Close door and run an empty Non Lumen test cycle.
3. If Unit alarms again, contact STERIS.
Door is fully closed.
1. Load must be repackaged and reprocessed after
Abort is completed.
2. Contact STERIS.
7.23 Door Open Too Long ALM
Condition
Door has been open for more than five minutes. Monitored at all times.
General Troubleshooting
Additional Information
Door is open.
1. Close door.
2. If alarm persists, contact STERIS.

Table of Contents

Related product manuals