3 — Installation Verification
17
10085896_H
V-PRO
®
maX 2
Operator Manual
3.3 Technical Specifications
This section includes the following topics:
Overall Size (W x L x H)
Operating Weight
Electric Requirements
Environmental Conditions
Sterilant
EMC
Equipment Drawings
Cycle Limits
Overall Size (W x L x H)
• 33 x 37-7/8 x 75-1/8” (838 x 962 x 1908 mm) Recessed
• 33 x 38-3/16 x 75-1/8” (838 x 973 x 1908 mm) Cabinet
Operating Weight
• Single Door
– 780 lb (354 kg) Recessed
– 920 lb (417 kg) Cabinet
• Double Door
– 910 lb (413 kg) Cabinet
Electric Requirements
• Three Phase (North America):
– 200-208/230 Vac, 50/60 Hz, 16 Amp
• Three Phase (Europe/Asia):
– 380-415 Vac, 50/60 Hz, 10 Amp
• Three Phase (Japan):
– 200 Vac, 50/60 Hz, 16 Amp
Environmental Conditions
• Temperature: 59 to 86°F (15 to 30°C)
• Relative Humidity Range: 20-65%
• A-Weighted Sound Power Level: 58 dBA (mean) - 83 dBA (maximum)
• Pollution Degree: 2
• Installation Category (Overvoltage Category): II
• Maximum Altitude: 6562’ (2000 m)
• IP20