3 — Installation Verification
15
10085896_H
V-PRO
®
maX 2
Operator Manual
Installation Verification
3.1 Introduction
An equipment drawing, showing all utility and space requirements, was supplied when the Low Temperature
Sterilization System was ordered. The clearance space specified on the equipment drawing is necessary for proper
operation of Unit. Uncrating and Installation Instructions (refer to Associated Publications of this manual for document
number) were furnished with the Unit. If documents are missing or misplaced, contact STERIS, giving the serial,
equipment and model numbers of the Unit. Replacement copies are sent promptly.
NOTE: Unit should not be used in an area which is not compatible with oxidizers.
3.2 Installation Checklist
Only STERIS or STERIS-trained personnel should install this Low Temperature Sterilization System. After the
Sterilization System is installed according to the provided instructions, complete the following checklist to ensure the
installation is complete and correct. Or, if so desired, contact STERIS to schedule an installation verification and/or a
demonstration of proper equipment operation.
This section includes the following:
Service Clearance
Services
Pre-Operation Checklist
Service Clearance
IMPORTANT: Insufficient clearance space makes repairs more difficult and time-consuming. Refer to the
equipment drawing for minimum clearance for service/maintenance access requirements.
◻ Clearance as specified on Equipment Drawing (see Associated Publications) must be available. Drawing was
previously furnished. If missing, contact STERIS for a replacement copy.
Services
◻ Electric disconnect must be clearly marked, provided with lockout capability, and located near the Sterilization
Unit.
◻ Electric service to the unit must be as specified on the machine data plate and equipment drawing.
◻ Electric service should be on a separate circuit, and not tied into circuits containing large reactive loads (e.g.,
motors).
Pre-Operation Checklist
DANGER
CHEMICAL INJURY HAZARD
When handling hydrogen peroxide, wear appropriate Personal Protective Equipment
(PPE; refer to Glossary) and observe all Safety Precautions. See Sterilant SDS, product
label and package insert for additional handling information.