7 Troubleshooting
77
10085896_H
V-PRO
®
maX 2
Operator Manual
7.67 Unit Does Not Alarm When Cup Is Empty NonALM
Condition
Unit does not notify user when Cup is empty (Non-Alarm).
General Troubleshooting Additional Information
Verify Cup is empty.
1. Select cycle. If Unit starts cycle, Cup is not empty.
2. If cycle does not start, contact STERIS.
7.68 Unit Not Calibrated ALM
Condition
Unit displays UNIT NOT CALIBRATED.
General Troubleshooting Additional Information
Control issue.
1. Contact STERIS.
System error.
1. Cycle power to Unit.
7.69 Vaporizer Overtemp ALM
Condition
Occurs if vaporizer temperature (RTD1) is above vaporizer over-temperature setpoint (>68 C: >154 F). Sterilizer
automatically aborts cycle.
General Troubleshooting Additional Information
Acknowledge alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
Check status screen.
1. If vaporizer is down to 60 C (140 F), reprocess load
and rerun cycle.
2. If temperature is above 68 C (154 F), contact
STERIS.