6 — Operation
35
10085896_H
V-PRO
®
maX 2
Operator Manual
6.5 Load Sterilization Unit
DANGER
FIRE AND EXPLOSION HAZARD
Liquid hydrogen peroxide is a strong oxidant and poses a FIRE, EXPLOSION OR
CONTAINER RUPTURE HAZARD. Avoid excessive heat, contamination or contact with
combustible materials. Clothing, shoes or other combustible materials that have come
in contact with hydrogen peroxide must be immediately and thoroughly washed with
water. Discard shoes contaminated with Sterilant in a fireproof container. If Sterilant is
allowed to dry in the materials, a fire may result. IN CASE OF FIRE, use water only.
CONTAIN SPILLS and dilute with 20 parts water. SEE Safety Data Sheet (SDS) for
additional information. EFFECTS MAY BE DELAYED AND A FIRE COULD START AT A
LATER TIME.
DANGER
PERSONAL INJURY, CONTAMINATED LOAD AND/OR EQUIPMENT DAMAGE HAZARD
● Processing a load containing a stainless-steel lumened instrument through the Non
Lumen or Fast Non Lumen (if equipped) Cycle results in an ineffective Sterilization
Cycle. Do not process stainless-steel lumened devices in a Non Lumen or Fast Non
Lumen Cycle.
● Failure to thoroughly clean, rinse and dry articles to be Sterilized could result in an
ineffective Sterilization Cycle.
WARNING
PERSONAL INJURY HAZARD
When closing the chamber door, keep hands and arms out of the door opening and
ensure opening is clear of any obstructions.
Load the Sterilization Unit as follows:
NOTE: STERIS recommends (in accordance with applicable ANSI/AAMI guidelines) wearing
chemical-resistant gloves when using the Sterilization Unit.
1. Pull chamber door open or step on the foot pedal.
2. Slide shelf half way out of Sterilization Unit chamber (see Figure 6-2).
3. Verify load is properly prepared (refer to Appendix A).