7.70 Vaporizer Temp Deviation ALM
Condition
In cycle, if RTD1 and RTD2 temperature readings differ by more than 0.5 C (1 F), then cycle aborts. When Unit is out of
cycle, alarm occurs if RTD1 and RTD2 temperature readings differ by more than 5 C (9 F).
General Troubleshooting
Additional Information
Acknowledge alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
7.71 Vaporizer Undertemp ALM
Condition
Occurs if vaporizer temperature (RTD1) is below vaporizer under-temperature setpoint (<59.5 C; <139 F) at any point
during a cycle. Sterilizer automatically aborts cycle.
General Troubleshooting Additional Information
Acknowledge alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
Check status screen.
1. If temperature is at least 59.5 C (139 F), reprocess
load and rerun cycle.
2. If temperature is below 59.5 C (139 F), contact
STERIS.
Room too cold.
1. Verify room temperature near unit is above 15 C or
59 F.