7 Troubleshooting
49
10085896_H
V-PRO
®
maX 2
Operator Manual
Troubleshooting
7.1 Introduction
IMPORTANT: A listing of the Safety Precautions to be observed when operating and servicing this equipment and/or handling
the Sterilant Cups can be found in Safety Precautions of this manual. Do not operate or service the equipment or handle the
Sterilant Cups until you have been trained on this information.
WARNING
PERSONAL INJURY AND/OR EQUIPMENT DAMAGE HAZARD
Repairs and adjustments to this equipment must be made only by STERIS or STERIS-trained service
personnel. Maintenance performed by unqualified personnel or installation of unauthorized parts could
cause personal injury, result in improper equipment performance, invalidate the warranty or result in
costly damage. Contact STERIS regarding service options.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment.
Contact STERIS to schedule preventive maintenance.
WARNING
ELECTRIC SHOCK HAZARD
Disconnect Sterilization Unit from facility electrical supply before servicing. Do not service this
Sterilization Unit unless electrical service has been properly locked out. Always follow appropriate
Lockout-Tagout and electrical safety-related work practice standards.
This section describes the possible alarm conditions/screens (see Typical Alarm Screen) the user may encounter while
operating this Low Temperature Sterilization System. See the individual alarms in this Section for a complete listing and
description of system alarms.
NOTE: When an alarm occurs the printer automatically generates a printout. Refer to Typical Alarm
Printouts Section.
If a problem occurs that cannot be corrected using this section or is not described in this section, please call STERIS. A
trained service technician will promptly place your Sterilization Unit in proper working condition.
NOTE: Never permit unqualified persons to service this Sterilization Unit.
Use the troubleshooting sections as follows:
Section Number and Alarm Name select the alarm that best describes the trouble symptom the Unit is
exhibiting.
Description Located directly below the alarm and describes what might have set it off.
Service Instructions follow the list of conditions that should be checked to isolate and correct the one causing
the problem. The instructions are listed in the order in which they should be completed.