Unload Sterilization Unit
DANGER
CHEMICAL INJURY HAZARD
Any visible liquids in the chamber or on the load must be treated as concentrated
hydrogen peroxide. Observe all hydrogen peroxide handling precautions. Refer to the
Sterilant Safety Data Sheet (SDS) for appropriate Personal Protective Equipment (PPE;
refer to Glossary), spill containment and cleanup.
The Sterilization Unit shelves slide out halfway to facilitate chamber unloading (see Figure 6-10). Ensure shelves are
returned to chamber before closing door.
ANSI/AAMI ST58, 2013, recommends using chemical-resistant gloves when using the Sterilization Unit.
NOTE: STERIS recommends leaving Sterilization Unit powered ON and door closed when not in use.
Unit needs approximately 45 minutes once SELECT CYCLE screen displays to reach operating
temperature if turned OFF.