7.62 Too Long to Inject — ALM
Condition
Occurs if time between start of injection (SV4 energizes) and inject complete (LS10 closes) exceeds eight seconds.
Sterilizer automatically aborts cycle.
NOTE:
Injection cylinder complete sensor (LS10) is a normally open switch that closes when injection cylinder is in up position.
General Troubleshooting
Additional Information
Acknowledge alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
Run empty Non Lumen cycle.
1. If it passes cycle, place back into service.
2. If it alarms again, contact STERIS.
7.63 Too Long to Inject Fill — ALM
Condition
Occurs if time between start of injection fill (SV8 energizes) and fill complete (LS11 closes) exceeds eight seconds.
Sterilizer automatically aborts cycle.
NOTE:
Injection cylinder fill sensor (LS11) is a normally open switch that closes when injection cylinder is in down position.
General Troubleshooting Additional Information
Acknowledge alarm.
1. Load must be repackaged and reprocessed after
Abort Phase is completed.
Run empty Non Lumen cycle.
1. If it passes, place back into service.
2. If it alarms again, contact STERIS.