3: Setting up PMUs and PGUs in Clarius  Model 4200A-SCS Pulse Card (PGU and PMU) 
 
3-28  4200A-PMU-900-01 Rev. B March 2023 
 
shown in the bottom graph of the following figure as six pulse periods. In this example, the pulse 
periods have voltage amplitude increasing from 0 V to 5 V. 
The graph labeled “Expanded View” provides in-depth information on the section designated by the 
cursor lines on the Entire Test graph. The graphed points reflect the timing parameters (Pulse Timing 
Preview) and settings from the PMU Advanced Settings dialog for each PMU channel in the test, but 
only for the area specified by the cursor. 
Figure 57: Six-point pulse amplitude sweep 
 
Select Preview Test to animate the waveform. The cursor moves from left to right on the bottom 
graph while displaying the point waveform of each sweep within the cursor on the top graph. 
Selecting Preview Test does not output any pulses, but it does illustrate how the test will progress 
as a result of the test parameter settings. 
PMU amplitude sweep and step example (two-channel) 
In this two-channel example, one channel is sweeping while the other channel is stepping. The 
sweeping channel is performing pulse amplitude sweeps with nine sweep points. On each inner loop,