Appendix A: Specifications
TDS 340A, TDS 360 & TDS 380 User Manual
A–5
Typical Characteristics
Typical characteristics are described in terms of typical or average performance.
Typical characteristics are not warranted.
Table A–6: Typical characteristics — signal acquisition system
Name Description
Accuracy, DC Gain, Envelope
Acquisition Mode
±3% for sec/div settings from 5 Sec/Div to 25 msec/div;
±2% for sec/div settings from 10 ms/div to 5 ns/div (TDS 340A);
±2% for sec/div settings from 10 ms/div to 2.5 ns/div (TDS 360);
±2% for sec/div settings from 10 ms/div to 1 ns/div (TDS 380)
Acc
u
r
a
c
y
,
DC V
olt
ag
e
M
e
a
s
u
reme
n
t,
S m le Ac isitio o e
Measurement type DC accuracy
S
a
m
p
le
Ac
qu
isitio
n M
o
d
e
Any Sample ±(2.0% × (|reading – Net Offset
1
|) + Offset
Accuracy + 0.13 div + 0.6 mV)
Delta Volts between any two samples
2
acquired under the same setup and
ambient conditions
±(2.0% × |reading| + 0.26 div + 1.2 mV)
Frequency Limit, Upper, 20 MHz Band-
width Limited
20 MHz
Ste
p R
es
p
o
n
se
Settli
ng E
rror
Vo
l
ts/D
iv
eig
St
ep
a pli e
Settling error (%)
3
s
e
tt
i
n
g
a
m
pli
tud
e
100 ns 20 ms
2 mV/div – 99.5 mV/div 2 V 1.0 0.1
100 mV/div – 995 mV/div 20 V 1.5 0.2
1 V/div – 10 V/div 200 V 2.5 0.2
Common Mode Rejection Ratio (CMRR) 100:1 at 60 Hz, reducing to 20:1 at 50 MHz, with equal Volts/Div and Coupling settings
on each channel.
1
Net Offset = Offset – (Position × Volts/Div). Net Offset is the voltage level at the center of the A-D converter dynamic
range. Offset Accuracy is the accuracy of this voltage level.
2
The samples must be acquired under the same setup and ambient conditions.
3
The values given are the maximum absolute difference between the value at the end of a specified time interval after the
mid-level crossing of the step, and the value one second after the mid-level crossing of the step, expressed as a
percentage of the step amplitude.

Table of Contents

Other manuals for Tektronix TDS 340A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 340A and is the answer not in the manual?

Tektronix TDS 340A Specifications

General IconGeneral
BrandTektronix
ModelTDS 340A
CategoryTest Equipment
LanguageEnglish

Related product manuals