EasyManuals Logo

Tektronix TDS 340A User Manual

Tektronix TDS 340A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
Appendix B: Performance Verification
TDS 340A, TDS 360 & TDS 380 User Manual
B–13
15. Press SET LEVEL TO 50%. Check that a stable trigger is obtained for the
test waveform on both the positive and negative slopes (see Figure B–7).
(Use the side menu to switch between trigger slopes; use the trigger LEVEL
knob to stabilize the trigger if required.)
First, set a signal with an
amplitude at the minimum trigger
sensitivity.
Second, check for a stable
trigger at both the positive and
negative slope settings.
Figure B–7: Measuring trigger sensitivity
16. Press WAVEFORM OFF.
17. Press CH 2.
18. Press TRIGGER MENU ! Source ! Ch2.
19. Disconnect the hookup from CH 1 and connect it to CH 2.
20. Set the vertical SCALE to 500 mV/div.
21. Repeat steps 14 and 15 for Channel 2.
22. Press TRIGGER MENU ! Source ! EXT/10.
23. Press MEASURE ! Select Measrmnt ! Amplitude.
24. Increase the generator amplitude until the amplitude measurement reads
1.5 V if you are checking a TDS 340A or TDS 360. Increase the generator
amplitude until the amplitude measurement reads 4.0 V if you are checking a
TDS 380.

Table of Contents

Other manuals for Tektronix TDS 340A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 340A and is the answer not in the manual?

Tektronix TDS 340A Specifications

General IconGeneral
BrandTektronix
ModelTDS 340A
CategoryTest Equipment
LanguageEnglish

Related product manuals