Appendix B: Performance Verification
B–8
TDS 340A, TDS 360 & TDS 380 User Manual
Equipment Required: One leveled sine wave generator (Item 7 or 8), one 50 W
precision cable (Item 2), and one 50 W termination (Item 1).
Time Required: Approximately 20 minutes.
Prerequisites: See page B–5.
Procedure:
1. Connect, through a 50 W precision cable and a 50 W termination, the sine
wave output of a leveled sine wave generator to CH 1 (see Figure B–3). Set
the output of the generator to a reference frequency of 50 kHz.
NOTE. If you are verifying a TDS 380, you need a leveled sine wave generator
with a 400 MHz output frequency. Refer to Sine Wave Generator Leveling
Procedure on page B–14 for information on obtaining a leveled output from an
unleveled sine wave generator.
Leveled
Sine Wave
Generator
Output
50 W Termination
Figure B–3: Hookup for analog bandwidth check
1. Press SAVE/RECALL SETUP ! Recall Factory Setup ! OK Confirm
Factory Init.
2. Set the horizontal SCALE to 10 ms/div.
3. Press TRIGGER MENU ! Coupling ! Noise Rej.
4. Press ACQUIRE ! Mode ! Average 16.
5. Press MEASURE ! High–Low Setup ! Min–Max.
6. Press the main menu button Select Measurement. Now press the side menu
button more until the menu label Pk-Pk appears in the side menu. Press the
side menu button Pk-Pk.
7. Set the vertical SCALE to 10 mV/div.
Check Analog Bandwidth

Table of Contents

Other manuals for Tektronix TDS 340A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 340A and is the answer not in the manual?

Tektronix TDS 340A Specifications

General IconGeneral
BrandTektronix
ModelTDS 340A
CategoryTest Equipment
LanguageEnglish

Related product manuals