EasyManua.ls Logo

Tektronix TDS 340A

Tektronix TDS 340A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Basics
2–14
TDS 340A, TDS 360 & TDS 380 User Manual
3. Check that the waveform indicates correct compensation (see Figure 2–6). If
the waveform indicates over or undercompensation, use the alignment tool
provided with the probe to adjust the compensation as shown in Figure 2–7.
Probe compensated correctly
Probe overcompensated
Probe undercompensated
Figure 2–6: How probe compensation affects signals
Compensation adjustment
Figure 2–7: Location of probe compensation adjustment

Table of Contents

Other manuals for Tektronix TDS 340A

Related product manuals