Commands
TDS5000B Series Service Manual
5-17
! TEMPDRIFT indicating that the instrument has not reached a state of st able
temperature adequate for performing signal path calibration
! RUNNING indicating that the signal path calibration is currently running
CALIBR ATE:INTERNal:STATus?
This query mi ght return INIT, indicating that the internal signal path calibration
has not been run.
CALibrate:PRObestate:CH<x>?
This query-only command returns the status of the probe calibration for a probe
attached to a given channel.
CH<x>:PROBECal?
CALibrate:PROBEstate:CH<x>?
<x> is the channel number the probe is attached to
--1 indicating that the probe calibration failed for the selected channel
0 indicating t hat the probe calibration passed for the selected channel
1 indicating t hat the probe calibration is ini t i alized for the selected channel
2 indicating t hat probe calibration is running for the selected channel
CALIBRATE:PROBESTATE:CH1?
This query returns the st atus of the probe calibration for a probe attached to
CH1.
A query might return
:CALIBRATE:PROBESTATE:CH1 0
indicating that the probe calibration for channel 1 passed.
CALibrate:RESults?
This query-only command returns the status of internal and factory calibration
without performing calibration operations. The returned results do not include
the calibration st atus of attached probes. The CALibration:RESults? query is
intended to support GO/NoGO testing of the instrument calibration readiness: all
returned results should indicate PASS status if the instrument is fit for duty. It is
Example
Related Commands
Syntax
Arguments
Returns
Example