Commands
5-16
TDS5000B Series Service Manual
CALibrate:INTERNal
This command st arts the internal signal path calibration (SP C) of the instrument.
You can use the CALibrate:INTERNal:STATus? query to return the current
status of the internal signal path calibration of the instrument.
CALibrate:INTERNal
CALIBRATE:INTERNal
This command st arts the internal signal path calibration of the instrument.
CALibrate:INTERNal:STARt
This command st arts the internal signal path calibration (SP C) of the instrument.
This command i s the same as the CALibrate:INTERNal command. You can use
the CALi brate:INTERNal:STATus? query t o return the current status of the
internal signal path calibration of the inst rument.
CALibrate:INTERNal:STARt
CALIBRATE:INTERNal:STARt
This command st arts the internal signal path calibration of the instrument.
CALibrate:INTERNal:STATus?
This query-only command returns the current status of the instrument internal
signal path calibration for the last SPC operation.
CALibrate:INTERNal:STATus?
This query wil l return one of the following:
! INIT indicating that the instrument has not had the internal signal path
calibration run
! PASS indicating that the signal path calibration completed successfully
! FAIL indicating that the signal path calibration did not complete successfully
! WARMUP indicating that the instrument has not warmed up for the required
time to perform signal path calibration
Syntax
Example
Syntax
Example
Syntax
Returns