Commands
5-20
TDS5000B Series Service Manual
Table 5- 5: D IAgnostic Commands
Command Des c ripti on
DIAg:STOP Termi nat es the execut ion of diagnosti cs
TEST Selects and executes any i tem at any level of the
test hierarchy
TEST:RESults? Returns a brief pass or fail status of the last test
execution
TEST:RESults:VERBos e ? Returns a mo re explanatory message about the
results of the last test execution
TEST:STOP Termi nates the execut io n of the test
DIAg:CONTROL:HALT
This command (no query form) determines whether the next execution of
diagnostics will stop on the first diagnostic failure that occurs or will execute the
selected set of diagnostic functions. This command is equivalent to selecting
Instrument Diagnostics from the Utilities menu and then enabling Halt on Fail.
DIAg:CONTROL:LOOP
DIAg:CONTROL:HALT (ON|OFF|<NR1>)
ON enables the halt function, causing the execution of diagnostics to halt at the
first diagnostic failure that occurs.
OFF disables the halt function, allowing the instrument to execute the entire set
of diagnostics before halting, even if a diagnostic failure occurs.
<NR1> A 0 enables the halt function; any other number disables t he halt
function.
DIAg:CONTROL:HALT ON
This command enables the halt function, causing the execution of diagnostics to
halt at the first diagnostic failure that occurs.
Related Commands
Syntax
Arguments
Example