Commands
TDS5000B Series Service Manual
5-21
DIAg:CONTROL:LOOP
This command (no query form) determines whether the next execution of
diagnostics executes once or continuously loops on the selected set of diagnos-
tics (assuming the halt control is set to off using the DIAg:CONTROL:HALT
command or that t he halt control is set to ON but no failures occur). This
command is equivalent to selecting Instrument Diagnostics from the Utilities
menu and then enabling Loop Control.
DIAg:CONTROL:HALT, DIAg:STOP
DIAg:CONTROL:LOOP (ON|OFF|<NR1>)
ON enables the loop function, causing the execution of diagnostics to continu-
ously loop.
OFF disables the loop function, causing the oscilloscope to execute t he entire set
of diagnostics once and then halt.
<NR1>--A 0 enables the loop function; any other number disables the loop
function.
DIAg:CONTROL:LOOP ON
This command enables the loop function.
DIAg:EXECUTE
This command (no query form) starts the execution of the currently selected set
of diagnostics. T hi s command is equivalent to selecting Instrument Diagnostics
from the Utilities menu and then pressing Run.
DIAg:STATE
DIAg:EXECUTE
DIAg:EXECUTE
This command st arts the execution of the selected set of diagnostics.
Related Commands
Syntax
Arguments
Examples
Related Commands
Syntax
Example